Evaluation of instrumentation used for high performance thin-layer chromatography of aflatoxins
Coker, R.D., Jewers, K., Tomlins, K.I. and Blunden, G. (1988) Evaluation of instrumentation used for high performance thin-layer chromatography of aflatoxins. Chromatographia, 25 (10). pp. 875-880. ISSN 0009-5893 (Print), 1612-1112 (Online) (doi:https://doi.org/10.1007/BF02311421)
Full text not available from this repository.Abstract
Instrumentation used for high performance thin layer chromatography (HPTLC) of aflatoxins has been evaluated. Optimum sensitivity, accuracy and precision resulted when a combination of a fully automated TLC sampler, an unsaturated conventional TLC glass chamber and a monochromatic densitometer were employed. Benzene: acetonitrile (98: 2 v/v) was found to be the most suitable sample application solvent. These findings are being used to develop a rapid, simple and reproducible method for the analysis of aflatoxins.
Item Type: | Article |
---|---|
Additional Information: | [1] Part 1 of a series “Development of rapid methods for the analysis of aflatoxin in agricultural produce”. |
Uncontrolled Keywords: | thin layer chromatography, sample application, densitometry, development chambers, aflatoxins |
Subjects: | Q Science > QD Chemistry S Agriculture > S Agriculture (General) |
Faculty / School / Research Centre / Research Group: | Faculty of Engineering & Science > Natural Resources Institute |