Skip navigation

Evaluation of instrumentation used for high performance thin-layer chromatography of aflatoxins

Evaluation of instrumentation used for high performance thin-layer chromatography of aflatoxins

Coker, R.D., Jewers, K., Tomlins, K.I. and Blunden, G. (1988) Evaluation of instrumentation used for high performance thin-layer chromatography of aflatoxins. Chromatographia, 25 (10). pp. 875-880. ISSN 0009-5893 (Print), 1612-1112 (Online) (doi:https://doi.org/10.1007/BF02311421)

Full text not available from this repository.

Abstract

Instrumentation used for high performance thin layer chromatography (HPTLC) of aflatoxins has been evaluated. Optimum sensitivity, accuracy and precision resulted when a combination of a fully automated TLC sampler, an unsaturated conventional TLC glass chamber and a monochromatic densitometer were employed. Benzene: acetonitrile (98: 2 v/v) was found to be the most suitable sample application solvent. These findings are being used to develop a rapid, simple and reproducible method for the analysis of aflatoxins.

Item Type: Article
Additional Information: [1] Part 1 of a series “Development of rapid methods for the analysis of aflatoxin in agricultural produce”.
Uncontrolled Keywords: thin layer chromatography, sample application, densitometry, development chambers, aflatoxins
Subjects: Q Science > QD Chemistry
S Agriculture > S Agriculture (General)
Faculty / School / Research Centre / Research Group: Faculty of Engineering & Science > Natural Resources Institute
Related URLs:
Last Modified: 06 Aug 2014 14:35
URI: http://gala.gre.ac.uk/id/eprint/11915

Actions (login required)

View Item View Item